Abstract

To improve the detection limits of a portable total reflection X-ray fluorescence (TXRF) spectrometer using white X-rays (i.e., both characteristic X-rays and continuum X-rays) from a 5 W X-ray tube, the measurement was performed in vacuum. The TXRF spectrum measured in vacuum was compared with that measured in air. The spectral background was significantly reduced when the scattering of the incident X-rays from air was reduced using a vacuum pump, leading to improvement in the detection limit. A detection limit of 8 pg was achieved for Cr when measuring in vacuum.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.