Abstract

We report on our efforts to measure simultaneously a well-calibrated complex impedance of a large number of detectors in a long wavelength bolometer array. The array is described in other presentations. A method for correcting the complex impedance measurements of bolometers and calorimeters has been presented by Lindeman et al. (Rev. Sci. Instrum. 78:043105, [2007]) using a Thevenin equivalent circuit to represent the bias network. We have built on this method for superconducting bolometers with a Norton equivalent circuit and have used it to improve our impedance data. We further describe our method for extracting a Norton-corrected complex impedance as a function of frequency from a stream of multiplexed time-ordered data. This method is well-suited to producing simultaneous complex impedance measurements for a large number of detectors.

Full Text
Published version (Free)