Abstract
This paper describes an electrocardiograph (ECG) monitoring SoC using a non-volatile MCU (NVMCU) and a noise-tolerant instantaneous heartbeat detector. The novelty of this work is the combination of the non-volatile MCU for normally off computing and a noise-tolerant-QRS (heartbeat) detector to achieve both low-power and noise tolerance. To minimize the stand-by current of MCU, a non-volatile flip-flop and a 6T-4C NVRAM are used. Proposed plate-line charge-share and bit-line non-precharge techniques also contribute to mitigate the active power overhead of 6T-4C NVRAM. The proposed accurate heartbeat detector uses coarse-fine autocorrelation and a template matching technique. Accurate heartbeat detection also contributes system-level power reduction because the active ratio of ADC and digital block can be reduced using heartbeat prediction. Measurement results show that the fully integrated ECG-SoC consumes 6.14 μ A including 1.28- μA non-volatile MCU and 0.7- μA heartbeat detector.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE transactions on biomedical circuits and systems
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.