Abstract
To evaluate the normalized average glandular dose (the average glandular dose per unit entrance skin exposure) in magnification mammography. Photon transport in the breast was simulated by using Monte Carlo methods. A semielliptical cylinder containing glandular and adipose tissue was used to simulate the breast. Measured mammography spectra for a molybdenum target-molybdenum filter unit were utilized. The normalized average glandular dose was calculated as a function of half-value layer, tube voltage, breast thickness, and breast composition for typical magnification geometries. The normalized average glandular dose in magnification mammography is 7%-25% lower than that with the contact (nonmagnification) technique because of the effects of partial irradiation, smaller field size, and greater percentage depth dose gradient at the reduced source-to-skin distance. The normalized average glandular dose in magnification mammography is lower than that in contact mammography. The average glandular dose in magnification mammography, however, is still substantially greater due to the two to three times greater entrance skin exposure.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.