Abstract

Mass spectrometry (MS) is an effective analytical tool for high‐throughput screening (HTS) in the drug discovery field. Infrared matrix‐assisted laser desorption electrospray ionization (IR‐MALDESI) MS is a high‐throughput platform that has achieved analysis times of sub‐seconds‐per‐sample. Due to the high‐throughput analysis speed, methods are needed to increase the analyte signal while decreasing the variability in IR‐MALDESI‐MS analyses to improve data quality and reduce false‐positive hits. The Z‐factor is used as a statistic of assay quality that can be improved by reducing the variation of target ion abundances or increasing signal. Herein we report optimal solvent compositions for increasing measured analyte abundances with direct analysis by IR‐MALDESI‐MS. We also evaluate normalization strategies, such as adding a normalization standard that is similar or dissimilar in structure to the model target drug, to reduce the variability of measured analyte abundances with direct analyses by IR‐MALDESI‐MS in both positive and negative ionization modes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.