Abstract

The authors report a case of nontraumatic, spontaneous dislocation of a polyethylene insert detected 1 year after total knee arthroplasty. The patient demonstrated initial improvement and returned to work 4 months postoperatively. At 6 months postoperatively, the patient developed pain and a clunking sensation with motion; however, he denied any traumatic precipitating events. An arthroscopic procedure revealed arthrofibrotic formations but no signs of locking mechanism failure. At 12 months postoperatively, the patient developed sudden instability, and radiographs demonstrated an anteriorly dislodged insert. Revision surgery was performed, and the insert was removed. The insert showed some signs of fatigue due to the locking mechanism. We postulated that repetitive flexion produced an anterior superior force leading to failure of the locking mechanism.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.