Abstract

Electromagnetic field (EMF) measurements, for labor safety and environment protection purposes, are performed in the near-field. Inaccuracy of the far-field EMF measurements oscillates around ±1 dB or bit worse; in the near-field measurements, errors at the level of ±6 dB must be sometimes accepted. In the case of non-stationary EMF measurements, their sense may be changed from quantitative to qualitative ones. In order to make it possible an estimation of the non-stationary EMF measurement accuracy, it is proposed new method of the meters calibration. The method is based upon a standard excitation with identical signal as the measured one. A set for the purpose includes a pulse generator (of frequency and pulse rating identical with the radiation source) and computer-controlled amplitude modulator that reflects the radiation pattern of an antenna (e.g. radar one) and its rotations. Contrary to calibration using monochromatic continuous wave (CW) excitation, proposed method should be repeated for any radiation source. The disadvantage is compensated by a possibility to estimate (additional) error of the method. Calculations show it at the level of 10%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.