Abstract
Calibration graphs for Laser-Excited Atomic Fluorescence Spectrometry (LEAFS) were constructed for graphite-tube electrothermal atomizers (ETAs). Laser radiation was propagated through holes in the side of the tubes, and atomic fluorescence was detected at a right angle through the bore of the tube. The LEAFS calibration graphs, while linear with a relative slope of one for 4 to 6 orders of magnitude, became nonlinear at high analyte concentrations. Observations concerning this nonlinearity are presented here. The onset of nonlinearity was believed to be primarily caused by post-filter effects. For nonresonance LEAFS, long lifetimes for the terminating energy levels in the fluorescence processes were believed to increase the probability of post-filter effects and subsequently to increase curvature of calibration graphs. Those conditions that increased the residence time of atoms in the furnaces (i.e., no internal gas flow and platform atomization) appeared to also increase post-filter effects and thus increased the amount of curvature of the calibration graphs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.