Abstract
In this paper, we report on nonlinearity compensation for a solid-state fiber bragg grating (FBG) sensor interrogation system based on an arrayed waveguide grating (AWG) device. A lookup table with calibration data is used to improve system linearity. A reduction in the absolute value of the measurement error from 120 mustrain or 15degC to 4.8 mustrain or 0.6degC, respectively, is experimentally demonstrated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.