Abstract
Abstract We measured the linearity of a Xe-filled gas proportional counter in the energy range between 10 and 40 keV at the synchrotron radiation facility. A clear gap at the Xe K-edge energy (34.5 keV) can be seen. We found that the equivalent width of this gap is about 170 ± 10 eV. In order to understand this value, we discussed the mean ionization energy during the X-ray photo absorption process.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.