Abstract

The noninvasive thermometry method is based on the temperature dependence of the proton resonance frequency (PRF). High-quality temperature images can be obtained from phase information of standard gradient-echo sequences with an accuracy of 0.2 degrees C in phantoms. This work was focused on the in vivo capabilities of this method. An experimental setup was designed that allows a qualitative in vivo verification. The lower-leg muscles of a volunteer were cooled and afterwards reheated with an external water bolus. The temperature of the bolus water varied between 17 degrees C and 37 degrees C. The in vivo temperature images can be used to extract the temperature in muscle tissue. The data in the fat tissue are difficult to interpret because of the predominance of susceptibility effects. The results confirm the method's potential for hyperthermia control.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.