Abstract

We have developed a noninvasive pneumatics-based system by which to measure heartbeat, respiration, snoring, and body movements of a subject in bed. A thin, air-sealed cushion is placed under the bed mattress of the subject and the small movements attributable to human automatic vital functions are measured as changes in pressure using a pressure sensor having an almost flat frequency response from 0.1 to 5 kHz and a sensitivity of 56 mV/Pa. Using the newly developed system, heartbeat, respiration, apnea, snoring and body movements are clearly measured. In addition, the optimal signal-to-noise (S/N) ratio by which to evaluate the reliability of the heart rate measurement is presented. Heart rates were measured for four different body postures, 13 different subjects, four different bed mattresses, and three different sensor positions. For these measurements, the S/N ratios ranged from 15.9 to 23.5 dB, and so were determined to be reliable.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.