Abstract

BackgroundThere may be significant difference between measurement of end-tidal carbon dioxide partial pressure (PetCO2) and arterial carbon dioxide partial pressure (PaCO2) during one-lung ventilation with low tidal volume for thoracic surgeries. Transcutaneous carbon dioxide partial pressure (PtcCO2) monitoring can be used continuously to evaluate PaCO2 in a noninvasive fashion. In this study, we compared the accuracy between PetCO2 and PtcCO2 in predicting PaCO2 during prolonged one-lung ventilation with low tidal volume for thoracic surgeries.MethodsEighteen adult patients who underwent thoracic surgeries with one-lung ventilation longer than two hours were included in this study. Their PetCO2, PtcCO2, and PaCO2 values were collected at five time points before and during one-lung ventilation. Agreement among measures was evaluated by Bland-Altman analysis.ResultsNinety sample sets were obtained. The bias and precision when PtcCO2 and PaCO2 were compared were 4.1 ± 6.5 mmHg during two-lung ventilation and 2.9 ± 6.1 mmHg during one-lung ventilation. Those when PetCO2 and PaCO2 were compared were -11.8 ± 6.4 mmHg during two-lung ventilation and -11.8 ± 4.9 mmHg during one-lung ventilation. The differences between PtcCO2 and PaCO2 were significantly lower than those between PetCO2 and PaCO2 at all five time-points (p < 0.05).ConclusionsPtcCO2 monitoring was more accurate for predicting PaCO2 levels during prolonged one-lung ventilation with low tidal volume for patients undergoing thoracic surgeries.

Highlights

  • Arterial carbon dioxide partial pressure (PaCO2) is the gold standard in monitoring ventilation during general anesthesia

  • We compared the accuracy between PetCO2 and PtcCO2 in predicting PaCO2 during prolonged one-lung ventilation with low tidal volume for thoracic surgeries

  • PtcCO2 monitoring was more accurate for predicting PaCO2 levels during prolonged onelung ventilation with low tidal volume for patients undergoing thoracic surgeries

Read more

Summary

Background

There may be significant difference between measurement of end-tidal carbon dioxide partial pressure (PetCO2) and arterial carbon dioxide partial pressure (PaCO2) during one-lung ventilation with low tidal volume for thoracic surgeries. Transcutaneous carbon dioxide partial pressure (PtcCO2) monitoring can be used continuously to evaluate PaCO2 in a noninvasive fashion. We compared the accuracy between PetCO2 and PtcCO2 in predicting PaCO2 during prolonged one-lung ventilation with low tidal volume for thoracic surgeries

Methods
Results
Conclusions
Introduction
Materials and Methods
Discussion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.