Abstract

ABSTRACT Purpose: Anterior chamber (AC) flare is a key sign for anterior uveitis. New instrument-based techniques for measuring AC flare can offer automation and objectivity. This review aims to identify objective instrument-based measures for AC flare. Methods: In this systematic review, we identified studies reporting correlation between instrument-based tests versus clinician AC flare grading, and/or aqueous protein concentration, as well as test reliability. Results: Four index tests were identified in 11 studies: laser-flare photometry (LFP), optical coherence tomography, ocular flare analysis meter (OFAM) and the double-pass technique. The correlation between LFP and clinician grading was 0.40–0.93 and 0.87–0.94 for LFP and protein concentration. The double-pass technique showed no correlation with clinician grading and insufficient information was available for OFAM. Conclusion: LFP shows moderate to strong correlation with clinician grading and aqueous protein concentration. LFP could be a superior reference test compared to clinician AC flare grading for validating new index tests.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.