Abstract

ObjectiveThe Alertgy noninvasive continuous glucose monitor (ANICGM) is a novel wristband device that reports glucose levels without entailing skin puncture. This study evaluated the performance of the ANICGM compared to a Food and Drug Administration-approved glucose meter in patients with type 2 diabetes. MethodsThe ANICGM device measures changes in the electromagnetic field generated by its sensor to produce a dielectric spectrum. The data contained within this spectrum are used in tandem with machine learning algorithms to estimate blood glucose levels. Values from the ANICGM were collected, sent to the Alertgy lab, formatted, and compared with fingerstick blood glucose levels, which were measured using the Accuchek Inform II glucometer. Fifteen patients completed three 120-minute sessions. The mean absolute relative difference (MARD) was calculated. ResultsMARD values were compared between study days 2 and 3. The MARD for day 2 was 18.5% (95% CI, 12.8-42.2%), and the MARD for day 3 was 15.3% (95% CI, 12.3-18.4%). The difference in the MARD between days 2 and 3 was not statistically significant (P = .210). ConclusionThe resulting MARDs suggest that further investigation into the use of dielectric spectroscopy for glucose monitoring should be explored.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.