Abstract

The present work demonstrates the feasibility of estimating blood glucose level using a microstrip antenna strapped on a patient's arm. Some promising success has been shown by the RIT ETA Lab research group that an antenna's resonant frequency can track, in real time, changes in glucose concentration. However, these changes in antenna response are unique to an individual. An equivalent circuit model has been developed to represent the input impedance with changing glucose levels. From a first set of measurements of input impedance in real time on a patient, the circuit elements are determined by curve fitting, in terms of the change in glucose level and incorporated into the expression for input impedance of the equivalent circuit The glucose level is estimated iteratively from subsequent measurements of the input impedance on the same patient The estimates, represented in the Clarke's Error Grid, compare well with the measurements made by a traditional blood glucose meter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.