Abstract

In this study, we developed an impedimetric sensor system by using molecularly cortisol imprinted acrylamide polymers on fullerene modified carbon electrode to detect cortisol in real saliva samples. The polymer layer was formed on fullerene modified screen printed carbon electrodes with the ratio of 1:2 (cortisol:acrylamide) and ammonium persulfate (APS) was used as the initiator of polymerization. The sensor surface was investigated by cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS) and Scanning Electron Microscopy (SEM). Performances of the electrode were determined as follows; calibration curve was calculated between 0.5 nM and 64 nM, R2 = 0.9939, LOD and LOQ were 0.14 nM and 0.44 nM, respectively. Real human saliva samples were obtained from 15 participants between 25 and 32 ages regardless of gender. The samples, which were collected at least 30 min after waking up and the new sensor method analysis were compared tandem mass spectroscopy coupled liquid chromatography system (LC-MS/MS). The sensor analysis showed a significant correlation with LC-MS/MS results (R2 = 0.9727).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.