Abstract

The Nexfin uses an uncalibrated pulse contour method for the continuous measurement of cardiac output (CO) in a totally noninvasive manner. Since the accuracy of pulse contour methods and their ability to track changes in CO have been repeatedly questioned, we have compared the CO measured by the Nexfin (NAPCO) with the CO measured by the pulmonary artery catheter (PACCO) in cardiosurgical patients before and after preload-modifying maneuvers. Twenty-eight patients who underwent on-pump cardiac surgery, of whom 18 were receiving vasopressor and/or inotropic therapy, were studied during the first postoperative hours. Preload modification, in the form of either a fluid challenge or a passive leg raising maneuver, was done whenever clinically indicated, with PACCO and NAPCO being simultaneously measured before and after each intervention. A fluid challenge was administered to 22 patients, and the passive leg raising maneuver was performed in 6 patients. These interventions were repeated in 19 patients producing a total of 47 pairs of measurements. At baseline, mean (±SD) CO was 4.9 ± 1.1 and 5.0 ± 1.4 L·min(-1), for the PACCO and NAPCO, respectively, bias 0.1 ± 1.0, 95% prediction interval -2.5 to 2.4 L·min(-1), and 39% of error. After preload modification, the mean CO was 5.6 ± 1.3 and 5.6± 1.5 L·min(-1) for the PACCO and NAPCO, respectively, bias -0.0 ± 1.1, 95% prediction interval -2.6 to 2.7 L·min(-1), and 38% of error. The correlation coefficients (r) between the PACCO and NAPCO before and after preload modification were 0.71 (95% confidence interval [95% CI], 0.53-0.82) and 0.70 (95% CI, 0.52-0.82), respectively. Preload modification induced similar absolute changes in PACCO and NAPCO (r = 0.9, P < 0.0001). A 4-quadrant scatter plot showed a concordance rate of 100% (95% CI, 80.5%-100%) between the changes in NAPCO and PACCO. Polar plot analysis demonstrated a small polar angle and radial limits of agreement well below the 30° benchmark. The area under a receiver operating characteristic curve, testing the ability of Nexfin to detect an increase of ≥15% in PACCO, was 0.974 (95% CI, 0.93-0.99). Although the Nexfin has limited accuracy when compared with the pulmonary artery catheter, it can reliably track preload-induced changes in CO in stable patients after cardiac surgery in the presence of moderate vasopressor and inotropic therapy. This ability, combined with its total noninvasiveness, fast installation, and ease of use, make the Nexfin a suitable monitor for the perioperative continuous measurement of CO. The reliability of this monitor in tracking the CO when significant changes in peripheral resistance take place still needs to be established.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.