Abstract

The purpose of this study was to examine the potential of using scatter information to evaluate tissue density at selected sites. A method for non-invasively generating profiles of density distribution within an object using Compton scattered X-rays is presented. The Compton scatter method is modified to scan longitudinal sections of composite phantoms and samples of tissue substitute materials. Line scan data are used to describe how the detected count rate changes in response to localized density variations within an extended object. The physical limitations of quantification are discussed, including the effect of attenuation, multiple scatter, and limited spatial resolution. Further, the theory of the method, its performance and results of experimental phantom studies are described. The results presented indicate that the suggested method has the potential for measuring physical density distribution within an object with a spatial resolution of 2 mm and an accuracy of approximately 4% under the circumstances in which attenuation correction is avoided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.