Abstract

The x-ray spectrum resulting from beta decay of an erbium tritide film is used to measure tritium content of the film. The areal density of the film is measured using differential emission techniques and is used to correct the measured x-ray intensity for self-absorption. The corrected intensity is linearly related to tritium content through an empirically derived calibration curve. The technique will measure tritium contents from less than one to several hundred micron liters. Repeatability is generally determined by counting errors and is typically 0.3% for a 5-min count while accuracy depends upon the empirical calibration technique. In principle, the technique is not limited to erbium tritide but can be extended to many uniformly radiating thin films.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.