Abstract

Angle‐resolved X‐ray photoelectron spectroscopy (ARXPS) has been used to characterize non‐destructively silicon oxynitride and high‐k film samples. The ARXPS data have been processed to provide accurate and precise measurements of thickness of surface and interface layers. Concentration depth profiles have been reconstructed from the ARXPS data to provide elemental and chemical state distribution information. For silicon oxynitride samples, nitrogen doses have been calculated from the concentration profiles, thereby accounting for the distribution of the nitrogen within the oxynitride layer. A comparison of ARXPS with modeled single‐angle XPS experiments illustrate the potential errors in calculation of both thickness and dose using the latter technique. Sputter depth profiles are also shown to contain potentially misleading information when compared to reconstructed ARXPS depth profiles.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.