Abstract

Background:There are very few studies on noncoplanar radiation in tomotherapy because deformable image registration is not implemented in the TomoTherapy Planning Station, a treatment planning device used in tomotherapy. This study examined whether noncoplanar radiation can be performed on the head using a tilt-type head and neck fixture and deformable image registration.Methods:Planning target volume spheres with diameters of 2, 3, and 4 cm were set on a head phantom, and computed tomography images were taken at 0° and 40° using a tilt-type head and neck fixture. Irradiation plans were created in the Tomotherapy Planning Station. Noncoplanar radiation was simulated, and the dose volume was evaluated by adding the 0° dose distribution and 40° dose distribution using the deformable image registration of the RayStation treatment planning system.Results:The ratio of the phantom volume to the irradiation dose for 20% to 30% of the planning target volume in noncoplanar radiation was smaller than that for 40% to 90% of the planning target volume in single-section irradiation at 0° or 40°.Conclusions:Noncoplanar radiation on the head region using tomotherapy was possible by using a tilt-type head and neck fixture, and the dose distribution could be evaluated using deformable image registration. This method helps reduce the dose of the organ-at-risk region located slightly away from the planning target volume.