Abstract

A new ultrasound instrument has been developed, using vector Doppler and embedded machine intelligence, to enable measurement of access flow rates by non-specialists. Ultrasound measurement of access flow can be performed with the patient off the dialysis machine, avoiding the hemodynamic changes that may affect indicator-dilution methods. A research version of the instrument was tested on flow phantoms simulating graft flow, and showed accuracy better than 5%. A non-specialist measured flow in the access grafts of 7 consenting dialysis patients; the instrument showed flows commensurate with indicator-dilution-measured flows, but with less variability. Measurements were made in less than 5 min per patient. The cost per measurement is calculated to be significantly less than that of present methods of measuring flow. The new instrument may become a useful tool for monitoring flow in accesses to extend their life.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.