Abstract

We demonstrate a new non-mydriatic ultra-widefield optical coherence tomography retinal imaging system, designed with custom optics to improve the imaging field of view, lateral resolution, and patient comfort. The key motivation is to address the challenge with conventional systems that require pupillary dilation, adding time, expense, discomfort, and medical risk to the examination of the retina. Our system provides an ultrawide 100° field of view (beam scanning angle at the scanning pivot point) and maintains a lateral resolution of 20 µm on the center. It also allows a generous working distance of 16 mm, 2-3 times longer than existing ultra-widefield OCT imaging systems. This advanced system was able to avoid iris vignetting artifacts without pharmacological dilation and ensure diffraction-limited ultra-widefield imaging under a generalized eye model. This enables a comprehensive evaluation of retina diseases, especially those affecting the peripheral regions.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.