Abstract
The Tensys T-line uses tonometry to reproduce the arterial blood pressure tracing non-invasively. The purpose of this study was to assess the agreement between estimates of the T-line and an intra-arterial catheter (for both mean arterial pressure [MAP] and pulse pressure variation [PPV]) in the setting of spine surgery. Continuous blood pressure data were collected for 7507 minutes from 25 patients. Five increasingly aggressive T-line filters were applied. The mean bias for mean, diastolic and systolic blood pressure ranged from 3.4–6.4, 3.1–7.1 and 0.1–0.8 mmHg and 6.5–11.8% for PPV. Ninety-five per cent confidence intervals for mean, diastolic and systolic blood pressure ranged from 24–28, 23.1–24.7 and 33.4–35.6 mmHg for 14–21% for PPV. The limits of agreement preclude the use of the T-line for reliable estimation of MAP or PPV in spine surgery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.