Abstract

BackgroundCardiac resynchronization therapy (CRT) is now generally delivered via quadripolar leads. Assessment of the effect of different vector programs from quadripolar leads on ventricular activation can be now done using non-invasive electrocardiographic mapping (ECM). Material and methodsIn nineteen patients with quadripolar LV leads, activation maps were constructed. The total ventricular activation time (TVaT) and the time for the bulk of ventricular activation (VaT10–90) were calculated. ResultsCRT delivered via a quadripolar lead significantly reduced TVaT and VaT10–90 by a mean of 16 ms and 31 ms, respectively, compared to baseline. There was a marked reduction in ventricular activation between the most and least synchronous vectors: 28% difference in baseline TVaT and 37% difference in VaT10–90. ConclusionChanges in the configuration of an LV quadripolar lead significantly affected ventricular activation timings in both ischaemic and non-ischaemic subjects. This suggests that programming of the optimal pacing vector may need to be individually tailored.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.