Abstract

Plethysmographic measurement of haemoglobin concentration ([Formula: see text]), pleth variability index (PVI), and perfusion index (PI) with the Radical-7 apparatus is growing in popularity. Previous studies have indicated that [Formula: see text] has poor precision, particularly when PI is low. We wanted to study the effects of a sympathetic block on these measurements. Twenty patients underwent hand surgery under brachial plexus block with one Radical-7 applied to each arm. Measurements were taken up to 20 min after the block had been initiated. Venous blood samples were also drawn from the non-blocked arm. During the last 10 min of the study, [Formula: see text] had increased by 8.6%. The PVI decreased by 54%, and PI increased by 188% in the blocked arm (median values). All these changes were statistically significant. In the non-blocked arm, these parameters did not change significantly. Brachial plexus block significantly altered [Formula: see text], PVI, and PI, which indicates that regional nervous control of the arm greatly affects plethysmographic measurements obtained by the Radical-7. After the brachial plexus block, [Formula: see text] increased and PVI decreased.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.