Abstract

The report describes a method of impedance cardiography using an improved estimate of thoracic volume. The formulas and their implementation in hardware and software are explained and new shortband electrodes are described which generate a good homogeneous thoracic field. Examples of stroke volume and cardiac output curves underline the capabilities of the monitoring system “Task Force ® Monitor”. In several experiments, results are compared to thermodilution as well as to BioZ measurements: the new method excels in comparison with thermodilution and is comparable to the BioZ device. Compared to traditional electrodes, the new shortband electrodes are shown to provide better reproducibility.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.