Abstract

In this paper, a sensor using a complementary split ring resonator (CSRR) is proposed for non-destructive testing of blood glucose. By depicting the complementary split ring structure on the ground, the electromagnetic field strength between the split rings can be enhanced effectively. The structure size of the sensor by CSRR is determined by simulation, so that the insertion loss curve of the device has a resonance point at the frequency of 3.419 GHz. With a special holder created by three-dimensional (3D) printing technology, the test platform was established when the concentration of the solution varied from 0 mg/mL to 20 mg/mL. The experimental results indicate that there is an obvious linear relationship between the insertion loss S21 and the glucose concentration at the resonant frequency. Similarly, the measured real part and imaginary part of the S21 both vary with glucose concentration linearly. Based on the above experimental results, the feasibility of the sensor using a CSRR proposed in this paper for non-destructive detection of blood glucose is preliminarily verified.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.