Abstract

Heart rate (HR) measurement is crucial for newborn infant monitoring in the neonatal intensive care unit (NICU). The widely used contact HR measurement methods based on electrocardiography (ECG) and photoplethysmography (PPG) signals can lead to discomfort and possible skin irritation on neonates, which limit its application in NICU scenarios. In this work, we propose a non-contact HR monitoring method simply using a RGB camera. Eulerian video magnification (EVM) is employed to detect the subtle changes of neonatal faces results from blood circulation. The magnified signal is then transformed to the spectral domain to extract HR information. Compared with the widely investigated independent component analysis (ICA)-based HR measurement method using video recordings, the proposed method can achieve the real time HR measurement, which is a significant superiority in NICU neonatal monitoring. To the best of our knowledge, this is the first study to employ EVM algorithm in real time neonatal HR monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.