Abstract

A THz camera based on an uncooled microbolometer 160X120 pixel array with nominal pitch of 52 μm has been developed at INO and initial transmission and reflection images showed promise. In the present paper, the characterization of both standard infrared and THz-optimized uncooled microbolometer pixel arrays are presented at both infrared and THz wavelengths. Measurements in the THz region has been performed using non-uniform low-power quantum-cascade laser (QCL) and uniform high-power far-infrared laser (FIR laser) beams at 3 THz and 4.25 and 2.54 THz, respectively. A measurement comparison has been achieved in the infrared using a blackbody radiation. Different methods for noise-equivalent power (NEP) measurements have been investigated. These characterization methods are promising especially for non-uniform laser beams irradiated on pixel arrays. The NEP results obtained from the different methods are in good agreement independent of the method used in the experiments. The results show a high sensitivity of the THz-optimized pixel array in the THz region. Large beam area reflection imaging of obscured materials at 2.54 THz have been performed at video rates of 30 frames per second using the THz-optimized pixel array equipped with a semi-custom fast THz objective, proving that the INO THz camera provides a promising solution for stand-alone imaging systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.