Abstract

AbstractLarge area digital imaging made possible by amorphous silicon thin-film transistor (a-Si TFT) technology, coupled with a-Si photo-sensors, provides an excellent readout platform to form an integrated medical image capture system. Major development challenges evolve aroundoptimization of pixel architecture for detector fill factor, signal propagation performance, and manufacturability, while suppressing noise stemming from pixel array and external electronics. This work analyzes a novel vertically integrated pixel design based on signal readout and noise performance, and compares with conventional co-planar and continuous detector architectures. In addition, the analysis will consider various substrate options including glass and robust substrates such as polymer and metal foil. Our evaluation have demonstrated state-of-the-art radiographic detector system with electronic noise under 2000 electrons at 150 µs frame time for an imaging arrays on robust substrate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.