Abstract
We present noise measurements of a phase fluorometric oxygen sensor that sets the limits of accuracy for this instrument. We analyze the phase sensitive detection measurement system with the signal ''shot'' noise being the only significant contribution to the system noise. Based on the modulated optical power received by the photomultiplier, the analysis predicts a noise spectral power density that was within 3 dB of the measured power spectral noise density. Our results demonstrate that at a received optical power of 20 fW the noise level was low enough to permit the detection of a change oxygen concentration of 1% at the sensor. We also present noise measurements of a new low-cost version of this instrument that uses a photodiode instead of a photomultiplier. These measurements show that the noise for this instrument was limited by noise generated in the preamplifier following the photodiode. (C) 1996 Society of Photo-Optical Instrumentation Engineers.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.