Abstract

Resolved NMR spectra from samples in inhomogeneous B 0 and B 1 fields can be obtained with the so-called “ex situ” methodology, employing a train of composite or adiabatic z-rotation RF pulses to periodically refocus the inhomogeneous broadening during the detection of the time-domain signal. Earlier schemes relied on a linear correlation between the inhomogeneous B 0 and B 1 fields. Here the pulse length, bandwidth, and amplitude of the adiabatic pulses of the hyperbolic secant type are adjusted to improve the refocusing for a setup with non-linear correlation. The field correlation is measured using a two-dimensional nutation experiment augmented with a third dimension with varying RF carrier frequency accounting for off-resonance effects. The pulse optimization is performed with a computer algorithm using the experimentally determined field correlation and a standard adiabatic z-rotation pulse as a starting point for the iterative optimization procedure. The shape of the z-rotation RF pulse is manipulated to provide refocusing for the conditions given by the sample-, magnet-, and RF-coil geometry.