Abstract

ObjectiveTo evaluate whether noninvasive-neurally adjusted ventilatory assist (NIV-NAVA) decrease respiratory efforts compared to nasal continuous positive airway pressure (NCPAP) during the first hours of life. MethodsTwenty infants born between 28+0 and 31+6 weeks were randomized to NIV-NAVA or NCPAP. Positive end-expiratory pressure was constantly kept at 6 cmH2O for both groups and the NAVA level was 1.0 cmH2O/µV for NIV-NAVA group. The electrical activity of diaphragm (Edi) were recorded for the first two hours. ResultsPeak and minimum Edi decreased similarly in both groups (P = 0.98 and P = 0.59, respectively). Leakages were higher in the NIV-NAVA group than in the NCPAP group (P < 0.001). The neural apnea defined as a flat Edi for ≥ 5 s were less frequent in NIV-NAVA group than in NCPAP group (P = 0.046). ConclusionsImmediately applied NIV-NAVA in premature infants did not reduce breathing effort, measured as peak Edi. However, NIV-NAVA decreased neural apneic episodes compared to NCPAP.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.