Abstract
PVC matrix and liquid membrane electrodes have been developed for direct potentiometric determination of gold(III). The membranes incorporate nitron tetrachloroaurate(III) as electroactive material. Fast response for gold(III) over the concentration range 10 −5–0.1 M, with response slopes of 52.8–55.2 mV/decade is obtained. The electrodes show good selectivity for gold(III) at pH 2–5 in the presence of many anions and cations. The PVC membrane electrode offers the advantages of greater selectivity (except for Cr 3+, Mn 2+ and ClO − 4) and higher thermal stability. The liquid membrane electrode gives a higher response slope and faster time of response than the PVC membrane electrode. Determination of AuCl − 4 over the range 2 μg/ml–2 mg/ml shows an average recovery of 98.5% and a mean standard deviation of 1.0%. Determination of gold in some gold alloys (58.3–99.9% Au) and pharmaceutical preparations gave an average recovery of 99.4% and a mean standard deviation of 0.7%, which are comparable with the performance obtained with the spectrophotometric Malachite Green and gravimetric U.S. Pharmacopeia methods.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.