Exogenic contamination is of primary concern for geochemical and biological clean laboratories working with sample sizes at the nanogram or even subpicogram level. Here, we determined sixty trace elements in fifteen different types of gloves from major suppliers world‐wide to evaluate whether gloves could be potential sources of contamination for routine trace element and isotope measurements. We found that all gloves contain some trace elements that can be easily mobilised in significant amounts. In weak acid at room temperature, the tested gloves released up to 17 mg of Zn, more than 1 μg of Mg, Ti, Mn, Fe, Rb, Sr, Zr, Sn, Hf and Pb and between 100 and 1000 ng of Li, Sc, V, Cr, Ni, Cu, Ga, As, Se, Y, Ag, Ba, La, Ce, Nd, Tl and Th. Vinyl gloves released lower quantities of biologically and geologically important elements, with the exception of In and Sn. Isotopic analyses indicate that all gloves share roughly the same Zn isotopic composition (average δ66Zn = +0.10 ± 0.32‰ (2s)). A single contact between glove and labware releases an average of ~ 6 ng of Zn and hence can significantly shift δ66Zn above the precision level when the amount of Zn determined is below 500 ng.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.