Abstract

We used a high-resolution mid-IR tunable-laser absorption spectroscopy (TLAS) system with a single IV-VI laser operating near 5.2 microm to measure the level of exhaled nitric oxide (eNO) in human breath. A method of internal calibration using simultaneous eNO and exhaled CO2 measurements eliminated the need for system calibration with gas standards. The results observed from internally calibrating the instrument for eNO measurements were compared with measurements of eNO calibrated to gas standards and were found to be similar. Various parameters of the TLAS system for eNO breath testing were examined and include gas cell pressure, exhalation time, and ambient NO concentrations. A reduction in eNO from elevated concentrations (approximately 44 parts in 10(9)) to near-normal levels (<20 parts in 10(9)) from an asthmatic patient was observed after the patient had received treatment with an inhaled glucocorticoid anti-inflammatory medication. Such measurements can help in evaluating airway inflammation and in monitoring the effectiveness of anti-inflammatory therapies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.