Abstract

We report the performance of a niobium hot-electron bolometer designed for laboratory terahertz spectroscopy. The antenna-coupled detector can operate above 4.2 K and has fast (subnanosecond) response. Detailed microwave measurements of performance over a wide range of operating conditions were correlated with quantitative terahertz measurements. The maximum responsivity is 4 x 10(4) VW with a noise equivalent power at the detector of 2 x 10(-14) W/Hz(12), approaching the intrinsic thermal fluctuation limit for the device. This detector enables a variety of novel laboratory spectroscopy measurements.