Abstract

The present work promotes and validates the benefits of using niobium instead of Havar® as the material for the target windows in most routine irradiations in cyclotrons. Calculation of the material activation and measurements of the contamination of the transferred target liquids show major improvements with the use of niobium. Also, the data of the daily routine productions at our production center are presented, proving that Havar® is not mandatory unless large target currents and/or pressures are required.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.