Abstract

Currently, positron emission tomography (PET) systems are rapidly developing owing to the invention of new scintillators and photosensors, which contribute to construction of clearer and high-resolution images in cancer diagnosis. In this study, we used a Lutetium Fine Silicate (LFS) inorganic scintillator and multi-pixel photon counter (MPPC) as a photosensor. This combination is realistic for next-generation PET systems. These devices were directly coupled to form a unit detector. A pair of detector units was placed face-to-face, and they simultaneously detected annihilation gamma rays, assuming a pair detector in the PET system. A timing resolution of 96 ps in full width at half maximum (FWHM) was obtained. This result is almost the fastest timing resolution ever achieved. A good timing resolution contributes to noise reduction that leads to clear images. Furthermore, a spatial resolution of 1.9 mm in FWHM was obtained, which is sufficient to meet the spatial resolution requirements for next-generation PET systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.