Abstract

A new leak element, which is named as “standard conductance element (SCE)”, has been developed for in situ calibration of ionization gauges (IGs) and quadrupole mass spectrometers (QMSs). The SCE is made of a stainless-steel sintered filter with the pore size of less than 1μm. Since the gas flow through the SCE satisfies the molecular flow condition even at the pressure up to 104Pa, some useful characteristics of molecular flow are available. The SCE is supplied to users with a calibration certificate described its molecular conductance. Users can introduce optional test gases with the known flow rate to their vacuum chamber through the SCE in their laboratories. The overview of the SCE, the calibration method and recommended practices are introduced.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.