Abstract

Background: While meaningful sound exposure has been shown to be important for newborn development, an excess of noise can delay the proper development of the auditory cortex. Aim: The aim of this study was to assess the acoustic environment of a preterm baby in an incubator on a newborn intensive care unit (NICU). Methods: An empty but running incubator (Giraffe Omnibed, GE Healthcare) was used to evaluate the incubator frequency response with 60 measurements. In addition, a full day and night period outside and inside the incubator at the NICU of the University Hospital Zurich was acoustically analyzed. Results: The fan construction inside the incubator generates noise in the frequency range of 1.3–1.5 kHz with a weighted sound pressure level (SPL) of 40.5 dB(A). The construction of the incubator narrows the transmitted frequency spectrum of sound entering the incubator to lower frequencies, but it does not attenuate transient noises such as alarms or opening and closing of cabinet doors substantially. Alarms, as generated by the monitors, the incubator, and additional devices, still pass to the newborn. Conclusions: The incubator does protect only insufficiently from noise coming from the NICUThe transmitted frequency spectrum is changed, limiting the impact of NICU noise on the neonate, but also limiting the neonate’s perception of voices. The incubator, in particular its fan, as well as alarms from patient monitors are major sources of noise. Further optimizations with regard to the sound exposure in the NICU, as well as studies on the role of the incubator as a source and modulator, are needed to meet the preterm infants’ multi-sensory needs.

Highlights

  • Introduction iationsNewborns born preterm require special care and spend days and sometimes even months in neonatal units equipped with incubators and monitoring tools supporting as well as controlling the infants’ cardiorespiratory function and temperature regulation

  • This study offers several approaches to reduce noise such as taking into account potential noise sources; we have not developed any concrete solution to this problem

  • 80 and 90 dB(A) due to the incubator opening or closing, which have been reviewed extensively [69,70]. These findings do not comply with the recommendations to maintain the combination of background and operational sound within an hourly equivalent continuous sound pressure level (SPL, Leq) of 50 dB, referring to a weighted slow response

Read more

Summary

Introduction

Newborns born preterm require special care and spend days and sometimes even months in neonatal units equipped with incubators and monitoring tools supporting as well as controlling the infants’ cardiorespiratory function and temperature regulation. These devices contain noise sources, including alarms, switches and fans. Results: The fan construction inside the incubator generates noise in the frequency range of 1.3–1.5 kHz with a weighted sound pressure level (SPL). Further optimizations with regard to the sound exposure in the NICU, as well as studies on the role of the incubator as a source and modulator, are needed to meet the preterm infants’

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.