Abstract

A turbidimetric assay was developed and validated against Escherichia coli for the quantitation of viable bacterial densities. The Abbott MS-2 research system was employed for continuous 5-min measurements of optical density. A linear standard curve was obtained by regressing the initial bacterial density (log CFU per milliliter) against the time required for bacterial growth causing a 5% decrease in optical transmittance. Slope and intercept values obtained from eight standard curves showed excellent assay reproducibility. Results obtained by the turbidimetric assay compared favorably to those obtained by the conventional pour plate assay. Prior to the application of the new assay, possible interferences of postantibiotic effect induced by the test antibiotics were excluded. The turbidimetric assay, which is presumably more efficient and less expensive, was implemented for the time-kill studies of three different beta-lactams against E. coli.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.