To perform stress-free recording of gastrointestinal motility in rats with strain gauge transducers, telemetry equipment had to be developed. We developed, programmed, and tested a new telemetry device that records gastrointestinal motility in freely moving rats using strain gauge transducers. The device can collect and transmit data in freely moving rats. Data are received and stored for later analysis with a regular PC. Linear calibration curves were obtained for the strain gauge transducers used. We compared data obtained with the new telemetry device with data gathered with standard equipment and could not find any statistically significant difference. Wired gastric and colonic contraction frequencies were 4.6 ± 0.3 per minute and 1.5 ± 0.3 per minute, whereas telemetric contraction frequencies were 4.4 ± 0.1 per minute and 1.25 ± 0.1 per minute. The new telemetry device is a very useful tool for the measurement of gastrointestinal motility in rats.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.