Abstract

A reference function for the post-processing uncertainty evaluation in specific absorption rate (SAR) compliance tests is proposed. The new reference function, derived from real SAR distributions measured during compliance sessions of handsets, is characterised by a double Gaussian distribution with asymmetrical standard deviation parameters, in each half space. Standard deviation parameters as well as SAR penetration are frequency dependent data. The proposed reference function also takes into account SAR distributions with two (or more) local maxima. The analysis for the derivation of the reference function parameters based on GSM900, DCS1800, UMTS and IEEE 802.11b measured SAR distributions is presented as well as details on uncertainty computations. A generalization of the method is presented too.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.