Abstract

A reverse flow injection spectrophotometric method for determination chlorine dioxide was developed with leucomethylene blue as the coloration reagent. By this method low concentrations of chlorine dioxide in water were determined in the presence of Cl2 and anion species such as OCl−, ClO2 − and ClO3 −. The linear range of ClO2 measurement is 0–1.91 mg/L with a detection limit of 0.02 mg/L. The presence of chlorine and hypochlorite ion can be masked by using oxalic acid. Chlorate does not interfere with the measurement. The permissible maximum concentration of ClO2 − is 3.0 mg/L.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.