Abstract

We have carried out a multicenter evaluation of a new reagent carrier for Reflotron, specific for pancreatic amylase where the salivary isoenzyme is inhibited by two specific monoclonal antibodies. This new procedure combines easy handling with low imprecision (median CV less than 3%) in control material, serum, heparinized blood, and plasma) and close correlation (r = 0.991 to 0.999) with established manual and automated methods. The same close correlation was found with values obtained from either venous or capillary finger-stick blood. Salivary amylase up to 54 kU/L (37 degrees C) was inhibited to about 97%. Endogenous interference by hemoglobin, bilirubin, triglycerides, cholesterol, or hematocrit was found to be negligible within a wide range of interferent concentrations. Out of a panel of 28 commonly used drugs it was shown that only two (ascorbic acid and paracetamol), and then only at toxic concentrations, caused a deviation in amylase activity of greater than 10%. From the results of this study we conclude that this new method is suitable for highly precise and accurate measurements of pancreatic amylase in emergency and routine laboratories.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.