Abstract

ABSTRACT This study reports the development of an all-in-one elemental analyser isotope ratio mass spectrometry (EA-IRMS) system modified for simultaneous analysis of dissolved organic carbon (DOC) concentration and its stable carbon isotope footprint (δ 13CDOC) in aqueous samples. The method involves a quantitative oxidation of DOC in a 200 µL liquid sample to CO2, after sample acidification and stripping by nitrogen. The detection limit of the method for DOC quantification was 0.2 mg C/L with an analytical precision of 12 %. Uncertainty of stable isotope determinations was 2 % at 0.2 mg DOC/L, while decreasing to 0.3 % at 20 mg DOC/L. Quantitative oxidation of DOC in aqueous samples was validated by using ring test water samples and Deep Sea reference seawater. The method performances of isotope analysis were evaluated by analysing different isotopic standard solutions. The applicability of the method was tested through the analysis of different environmental types of water, showing that δ 13CDOC ranged from – 23.30 to –31.85 ‰, allowing to characterize samples of different environmental origin. The developed method offers several advantages including rapidity, use of small sample volumes and minimal sample pre-treatment, making it a valuable tool for routine DOC concentration measurements paired with isotopic characterization.