Abstract
Nickel, a component of stainless steels (SS) applied in orthopedic implants may cause allergic processes in human tissues. P558 nickel free SS was studied to verify its viability as a substitute for stainless steel containing nickel. Its performance is compared to ISO 5832-9 and F138 most used nowadays grades in implants fabrications, in minimum essential medium, MEM, at 37 °C. Potentiodynamic polarization curves, electrochemical impedance spectroscopy (EIS), scanning electron microscopy (SEM) and “in vitro” cytotoxicity were used as techniques. From the electrochemical point of view P558 SS is comparable to ISO 5832-9 SS in MEM. It remains passivated until the transpassivation potential, above which generalized corrosion occurs. F138 presents pitting corrosion at 370 mV/SCE. The cytotoxicity results showed that P558, ISO 5832-9 and F138 do not present cytotoxic character. Therefore, these results suggest that P558 SS can be applied in orthopedic implants.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.